Regulations of Connecticut State Agencies (Last Updated: June 14,2023) |
Title19 Public Health and Safety |
SubTitle19-25d-1_19-25d-11. X-Ray Devices used for Diagnosis and Therapy |
Sec.19-25d-7. Special requirements for chest photofluorographic installations
Latest version.
-
(a) Equipment
(1) All provisions of subsection 19-25d-2 (a) apply.
(2) A collimator shall restrict the useful beam to the area of the photofluoro-graphic screen.
(b) Structural shielding. All provisions of subsections 19-25d-3 and 19-25d-5 (b) apply.
(c) Operating procedures
(1) All provisions of subsection 19-25d-5 (c) apply.
(2) All individuals except the patient being examined shall be in shielded positions during exposures.
(3) Personnel monitoring shall be required for all individuals operating the equipment.
(Effective October 1, 1982)