Sec.19-25d-7. Special requirements for chest photofluorographic installations  


Latest version.
  • (a) Equipment

    (1) All provisions of subsection 19-25d-2 (a) apply.

    (2) A collimator shall restrict the useful beam to the area of the photofluoro-graphic screen.

    (b) Structural shielding. All provisions of subsections 19-25d-3 and 19-25d-5 (b) apply.

    (c) Operating procedures

    (1) All provisions of subsection 19-25d-5 (c) apply.

    (2) All individuals except the patient being examined shall be in shielded positions during exposures.

    (3) Personnel monitoring shall be required for all individuals operating the equipment.

(Effective October 1, 1982)